Valleylab force 40
Description
Description:
The Valleylab Force 40 is a digital electrosurgical unit that displays power and is very flexible in terms of cautery and procedure capabilities. The Force 40 ESU will perform standard functions including blend for both cut and bipolar modes.
SPECIFICATIONS
General | Automatic activation impedance 40-100ohms Automatic deactivation impedance: <20ohms, >2000ohms Time delay range: .1 to 5.1 seconds Activation tone: 45 to 65 dBA |
Alarm Tone: >65 dBA Weight: 20 lbs (9.1 kg) Dimensions: 7.8 (H) × 16.1 (W) × 11.9 (D) inches (199 × 409 × 303 mm) Operates on 110/240V, 50/60Hz. |
Output Waveform | Cut: 500kHz sinusiod Blend 1: 500kHz bursts of sinusiod @ 50% duty cycle, recurring @ 31.25 kHz Blend 2: 500kHz bursts of sinusiod @ 37.5% duty cycle, recurring @ 31.25 kHz Desiccate: 500kHzdamped sinusiodal burst with a repetition frequency of 250 kHz |
Fulgurate: 500kHz damped sinusiodal burst with a repetition frequency of 62.5 kHz Spray: 500kHz damped sinusiodal burst with a repetition frequency of 31.25 kHz Standard Bipolar: 500kHz sinusiod Precise Bipolar: 500kHz sinusiod |
Output Characteristics | Cut: Max P-P 3300, Rate Load 300, Max Power 300, Crest Factor 2.1 @ 100W Blend 1: Max P-P 3800, Rate Load 300, Max Power 250, Crest Factor 2.1 @ 100W Blend 2: Max P-P 4000, Rate Load 300, Max Power 200, Crest Factor 2.1 @ 100W Desiccate: Max P-P 3500, Rate Load 300, Max Power 200, Crest Factor 2.1 @ 100W |
Fulgurate: Max P-P 6500, Rate Load 300, Max Power 150, Crest Factor 2.1 @ 100W Spray: Max P-P 9000, Rate Load 300, Max Power 150, Crest Factor 2.1 @ 100W Standard Bipolar: Max P-P 550, Rate Load 100, Max Power 50, Crest Factor 2.1 @ 100W Precise Bipolar: Max P-P 550, Rate Load 100, Max Power 50, Crest Factor 2.1 @ 100W |